| Model Number |
LSCD201 |
| Medical Device Problem Codes |
Break (1069); Activation Failure (3270); Premature Separation (4045)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/17/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the patient presented in clinic for an initial implant procedure of an aveir system on (b)(6) 2026.While positioning the atrial leadless pacemaker (alp), the physician had difficulties transitioning the delivery catheter (dc) into long tether mode and the alp prematurely separated from the dc due to repeated unsuccessful tether mode activation attempts.The alp dislodged and embolized to the pulmonary artery.The alp was retrieved and the physician elected to implant a ventricular device only.Assessment of the dc noted the tether may have been fractured within the dc itself due to being unable to move the release knob.The patient was stable throughout the procedure.
|
| |
|