| Model Number |
URF-V3 |
| Medical Device Problem Codes |
Loss of or Failure to Bond (1068); Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/02/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer returned the uretero-reno videoscope to the service center for a separate issue.During the device initial analysis, the service repair team noted that
the adhesive on the device a-rubber bending section was found to be peeled and needed to be replaced.There was no patient involvement in this reported event.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was evaluated.Evaluation confirmed the reported issue.A definitive root cause could not be identified.Reasonably known information suggests probable causes such as stress, damage or deterioration of parts , external factors.Olympus will continue to monitor field performance for this device.
|
| |
|