| Model Number |
LifeVest WCD 5100 System |
| Medical Device Problem Codes |
Structural Problem (2506); Key or Button Unresponsive/not Working (4063)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/02/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A us distributor returned a monitor and reported that the response buttons were non-functional.
|
| |
|
Event or Problem Description
|
|
A us distributor reported that an electrode belt was unable to securely latch to a monitor.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation of monitor has been completed.The reported problem (response buttons non-functional) was due to the response button switch was broken.The root cause of the broken switch cannot be positively identified.There was no adverse event that resulted from the damaged monitor.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation of electrode belt has been completed.The reported problem (unable to securely latch to a monitor) has been confirmed.Upon evaluation of the electrode belt, the trunk cable connector was damaged and was unable to connect to a monitor.The root cause for the damaged connector could not be positively identified.There was no adverse event that resulted from the damaged belt.
|
| |
|