| Model Number |
CYF-V2 |
| Medical Device Problem Codes |
Corroded (1131); Crack (1135)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/20/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer returned the videoscope for a separate issue.During the initial evaluation of the device, it was observed that the plastic distal end cover, the bending section cover and its adhesive, objective lens, light guide lens, and insertion tube boot were corroded.The distal end's plastic cover was also chipped.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
Based on the final investigation of the device the initial evaluation findings were confirmed.A root cause was unable to be identified.Olympus will continue to monitor the field performance of this device.Should any additional relevant information become available, a supplemental report will be submitted.
|
| |
|