The account reported finding incorrect items and leaking tubing.Three cases were received to aid in the investigation; the packs were opened and the tubing was reviewed, visual inspection noted no discrepancies, no cracks, holes or other defects were noted, and water was run through the tubing with no leaking noted at any point.Therefore, the defect reported could not be confirmed with the sample provided.Based on the review of the samples provided, medline was unable to replicate the reported issue of the injection tubing for the power injector leaking, and after reviewing the pack history was not able to determine what components were incorrect within the pack as no changes were noted between the revisions of the pack.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|