| Model Number |
2110 |
| Medical Device Problem Code |
Failure to Infuse (2340)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/27/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the device generated an "air in line" alarm during normal clinical use.This alarm condition is associated with a high-priority alarm and could potentially interrupt therapy if it occurs during patient use.Patient involvement is unknown; however, no patient harm or adverse event was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The suspect pump was returned for evaluation.The event history log (ehl) was reviewed, and multiple "air in line" alarms were identified.A review of the 12-month service history was conducted, and no relevant findings were identified.Visual inspection and functional testing were performed.The customer-reported issue was confirmed.The cause of the reported issue was determined to be a faulty air sensor.The air sensor was replaced, and the pump was restored to proper functionality.
|
| |
|