• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ICU Medical, Inc. CADD Solis HPCA Pumps; Pump, infusion, pca

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ICU Medical, Inc. CADD Solis HPCA Pumps; Pump, infusion, pca Back to Search Results
Model Number 2110
Medical Device Problem Code Failure to Infuse (2340)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/27/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the device generated an "air in line" alarm during normal clinical use.This alarm condition is associated with a high-priority alarm and could potentially interrupt therapy if it occurs during patient use.Patient involvement is unknown; however, no patient harm or adverse event was reported.
 
Additional Manufacturer Narrative
The suspect pump was returned for evaluation.The event history log (ehl) was reviewed, and multiple "air in line" alarms were identified.A review of the 12-month service history was conducted, and no relevant findings were identified.Visual inspection and functional testing were performed.The customer-reported issue was confirmed.The cause of the reported issue was determined to be a faulty air sensor.The air sensor was replaced, and the pump was restored to proper functionality.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CADD Solis HPCA Pumps
Common Device Name
Pump, infusion, pca
Manufacturer (Section D)
ICU Medical, Inc.
6000 nathan lane north
minneapolis MN 55442
Manufacturer Contact
reed covert
600 n. field dr.
lake forest, IL 60045
+1(224)706-2300
MDR Report Key50594286
Report Number3012307300-2026-08487
Device Sequence Number9451655
Product Code MEA
UDI-Device Identifier15019517099690
UDI-Public(01)15019517099690(11)190305
Combination Product (Y/N)N
Initial Reporter StateNA
Initial Reporter CountryAS
PMA/510(K) Number
K170982
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 09/03/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model Number2110
Device Catalogue Number21-2111-0401-51
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer08/07/2026
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date09/03/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/05/2019
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-