• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Abbott Medical (CRM-Sylmar) OPTISURE ACTIVE FIXATION, DF-4 CONNECTOR; Permanent Defibrillator Electrodes

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Abbott Medical (CRM-Sylmar) OPTISURE ACTIVE FIXATION, DF-4 CONNECTOR; Permanent Defibrillator Electrodes Back to Search Results
Model Number LDA210Q/65
Medical Device Problem Codes Failure to Capture (1081); Over-Sensing (1438); Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Serious Injury
Event or Problem Description
It was noted that the patient presented remotely via merlin.Net.Review of the transmission revealed that the right ventricular (rv) and right atrial (ra) leads exhibited loss of capture and oversensing.Fluoroscopy confirmed that the rv and ra leads were dislodged and located in the incorrect chambers.The leads were explanted and replaced on (b)(6) 2026.The patient remained in stable condition.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPTISURE ACTIVE FIXATION, DF-4 CONNECTOR
Common Device Name
Permanent Defibrillator Electrodes
Manufacturer (Section D)
Abbott Medical (CRM-Sylmar)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
Abbott Medical (CRM-Sylmar)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
navpreet singh
15900 valley view court
sylmar, CA 91342
MDR Report Key50594512
Report Number2017865-2026-18653
Device Sequence Number10067524
Product Code NVY
UDI-Device Identifier05414734507332
UDI-Public(01)05414734507332(10)A000187377(17)290430
Combination Product (Y/N)Y
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
P950022
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Device Explanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/03/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberLDA210Q/65
Device Lot NumberA000187377
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date09/03/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/06/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
Gallant DR
Outcome Attributed to Adverse Event Required Intervention;
Patient Age71 YR
Patient SexFemale
-
-