The customer observed false reactive alinity s hiv ag/ab combo and provided the following data:
sample id (b)(6) results are 1.84, 2.84, and 2.86.Nat: negative.Sample id (b)(6) results are 1.41, 1.43, and 1.49.Nat: negative.Sample id (b)(6) results are 1.07, 2.35, and non-reactive (result not provided).Nat: negative.Sample id (b)(6) results are 1.00, 6.66, and non-reactive (result not provided).Nat: negative.Sample id (b)(6) results are 15.65, 34.64, and 4.10.Nat: negative.Sample id (b)(6) results are 13.15, 18.05, and 30.90.Nat: negative.Sample id (b)(6) results are 4.04, 4.06, and 6.20.Nat: negative.Sample id (b)(6) results are 10.93, 19.58, and 20.23.Nat: negative.Sample id (b)(6) results are 37.48, 45.01, and 43.48.Nat: negative.No results were provided for sample ids (b)(6) and (b)(6).The customer reported that all samples generated negative nat results.The customer further communicated that the samples were for the purpose of tissue donation, however the customer confirmed discrepant results were not reported out to the medical provider.There was no reported impact to patient management.
|