• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Thoratec Corporation HEARTMATE®, MOBILE POWER UNIT, N/A; Ventricular (Assist) Bypass

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Thoratec Corporation HEARTMATE®, MOBILE POWER UNIT, N/A; Ventricular (Assist) Bypass Back to Search Results
Model Number 107754
Medical Device Problem Code Complete Loss of Power (4015)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/09/2026
Type of Reportable Event Malfunction
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEARTMATE®, MOBILE POWER UNIT, N/A
Common Device Name
Ventricular (Assist) Bypass
Manufacturer (Section D)
Thoratec Corporation
6035 stoneridge drive
pleasanton CA 94588
Manufacturer (Section G)
Thoratec Corporation
6035 stoneridge drive
pleasanton CA 94588
Manufacturer Contact
fredric gold
6035 stoneridge drive
pleasanton, CA 94588
+1(781)852-8207
MDR Report Key50597879
Report Number2916596-2026-2919324
Device Sequence Number15875747
Product Code DSQ
UDI-Device Identifier00813024010883
UDI-Public(01)00813024010883
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
P160054
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/24/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number107754
Device Catalogue Number107754
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Supplement Date Received by Manufacturer09/23/2026
Initial Report FDA Received Date09/03/2026
Supplement Report FDA Received Date09/24/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient Age56 YR
Patient SexMale
Patient Weight184 LB
-
-