BOSTON SCIENTIFIC CORPORATION FARAWAVE; See H11 Back to Search Results
Lot Number 0038834798
Medical Device Problem Code Material Twisted/Bent (2981)
Health Effect - Clinical Code Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 08/05/2026
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
D2a: common device name: percutaneous cardiac ablation catheter for treatment of atrial fibrillation with irreversible electroporation; reported here as the common device name exceeded the character limit for designated field.
 
Event or Problem Description
During a procedure, a farawave catheter was selected for use.Upon deployment after insertion into the patient, a spline inversion occurred and could not be corrected while the catheter was in situ.The catheter was therefore replaced, and the procedure was completed successfully with no complications reported during case.Following the procedure, the patient developed gastric distention and was hospitalized.The condition resolved by the following day, and the patient was subsequently discharged.At the time of reporting, the underlying cause of the gastric dilatation remained unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FARAWAVE
Common Device Name
See H11
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
Boston Scientific Corporation
302 parkway, global park
la aurora - heredia
CS  
Manufacturer Contact
mary crawford
4100 hamline avenue north
dc a330
saint paul, MN 55112
+1(651)440-0979
MDR Report Key50599638
Report Number2124215-2026-47197
Device Sequence Number21371445
Product Code QZI
UDI-Device Identifier00191506043216
UDI-Public(01)00191506043216(17)270314(10)38834798
Combination Product (Y/N)N
Initial Reporter CountryJA
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/17/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Lot Number0038834798
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Supplement Date Received by Manufacturer09/11/2026
Initial Report FDA Received Date09/03/2026
Supplement Report FDA Received Date09/17/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured03/14/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other; Hospitalization;