| Model Number |
M635WU60350 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Thrombosis/Thrombus (4440)
|
| Date of Event |
07/30/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
With all the available information, boston scientific (bsc) concludes:
device technical analysis
the watchman flx? pro remains implanted; therefore, returned device analysis could not be completed.Device history record review
a review was completed of the device history records (dhr) for the watchman flx? pro, part # m635wu60350 batch # 0036043535.The dhr review confirmed that the device met all material, assembly, and performance specifications prior to distribution.Labeling review
there was no evidence to suggest that the device was used in a manner inconsistent with the labeling.Watchman flx? pro instructions for use (ifu) lists potential complications, risks and adverse events that may be associated with the use of this device which include thrombosis.Risk review
a risk review was completed and confirmed that the event of thrombosis was defined in the risk documentation.This event type has been accounted for during product risk analysis to support acceptable risk benefit for the product.Investigation conclusion
based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of no problem detected.
|
| |
|
Event or Problem Description
|
|
It was reported that thrombosis occurred.A left atrial appendage (laa) closure procedure was performed, and a 35mm watchman flx pro closure device was implanted on (b)(6) 2026.The patient was discharged on eliquis and switched to dual antiplatelet therapy (dapt) at the routine 45 day follow up.The patient had a routine five (5) months follow up computed tomography (ct) and imaging showed a non-mobile 7x10mm lobulated device related thrombus (drt) located on the posterior margin of the closure device.The patient was then placed back on eliquis on (b)(6) 2026 and planned for a follow up transesophageal echocardiography (tee) in three (3) months.
|
| |
|
Event or Problem Description
|
|
It was reported that thrombosis occurred.A left atrial appendage (laa) closure procedure was performed, and a 35mm watchman flx pro closure device was implanted on (b)(6) 2026.The patient was discharged on eliquis and switched to dual antiplatelet therapy (dapt) at the routine 45 day follow up.The patient had a routine five (5) months follow up computed tomography (ct) and imaging showed a non-mobile 7x10mm lobulated device related thrombus (drt) located on the posterior margin of the closure device.The patient was then placed back on eliquis on (b)(6) 2026 and planned for a follow up transesophageal echocardiography (tee) in three (3) months.
|
| |
|