• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION WATCHMAN FLX? Pro; system, appendage closure, left atrial

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION WATCHMAN FLX? Pro; system, appendage closure, left atrial Back to Search Results
Model Number M635WU60350
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Thrombosis/Thrombus (4440)
Date of Event 07/30/2026
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
With all the available information, boston scientific (bsc) concludes: device technical analysis the watchman flx? pro remains implanted; therefore, returned device analysis could not be completed.Device history record review a review was completed of the device history records (dhr) for the watchman flx? pro, part # m635wu60350 batch # 0036043535.The dhr review confirmed that the device met all material, assembly, and performance specifications prior to distribution.Labeling review there was no evidence to suggest that the device was used in a manner inconsistent with the labeling.Watchman flx? pro instructions for use (ifu) lists potential complications, risks and adverse events that may be associated with the use of this device which include thrombosis.Risk review a risk review was completed and confirmed that the event of thrombosis was defined in the risk documentation.This event type has been accounted for during product risk analysis to support acceptable risk benefit for the product.Investigation conclusion based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of no problem detected.
 
Event or Problem Description
It was reported that thrombosis occurred.A left atrial appendage (laa) closure procedure was performed, and a 35mm watchman flx pro closure device was implanted on (b)(6) 2026.The patient was discharged on eliquis and switched to dual antiplatelet therapy (dapt) at the routine 45 day follow up.The patient had a routine five (5) months follow up computed tomography (ct) and imaging showed a non-mobile 7x10mm lobulated device related thrombus (drt) located on the posterior margin of the closure device.The patient was then placed back on eliquis on (b)(6) 2026 and planned for a follow up transesophageal echocardiography (tee) in three (3) months.
 
Event or Problem Description
It was reported that thrombosis occurred.A left atrial appendage (laa) closure procedure was performed, and a 35mm watchman flx pro closure device was implanted on (b)(6) 2026.The patient was discharged on eliquis and switched to dual antiplatelet therapy (dapt) at the routine 45 day follow up.The patient had a routine five (5) months follow up computed tomography (ct) and imaging showed a non-mobile 7x10mm lobulated device related thrombus (drt) located on the posterior margin of the closure device.The patient was then placed back on eliquis on (b)(6) 2026 and planned for a follow up transesophageal echocardiography (tee) in three (3) months.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WATCHMAN FLX? Pro
Common Device Name
system, appendage closure, left atrial
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
Boston Scientific Corporation
ballybrit business park
galway H91 Y 868
EI   H91 Y868
Manufacturer Contact
mary crawford
4100 hamline ave n
arden hills, MN 55112
+1(651)440-0979
MDR Report Key50600088
Report Number2124215-2026-45686
Device Sequence Number16108504
Product Code NGV
UDI-Device Identifier00191506004620
UDI-Public(01)00191506004620(17)280310(10)36043535
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
P130013
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/24/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberM635WU60350
Device Catalogue NumberM635WU60350
Device Lot Number0036043535
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Supplement Date Received by Manufacturer09/23/2026
Initial Report FDA Received Date09/03/2026
Supplement Report FDA Received Date09/24/2026
Date Device Manufactured03/11/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age80 YR
Patient SexMale
-
-