| Model Number |
D022 |
| Medical Device Problem Codes |
Pacing Problem (1439); Under-Sensing (1661)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/05/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that this implantable cardioverter defibrillator (icd) stored a ventricular episode where inappropriate anti-tachycardia pacing (atp) was delivered due to supraventricular tachycardia (svt).1:1 condition and ventricular sense beat (vs) were blanking out the atrial sense beat (as), which led to the delivery of therapy due to v>a.Technical services (ts) discussed troubleshooting options and programming considerations, and recommended turning off atrial tachy discriminators with rhythmid.This icd remains in service.No adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
If pertinent information is provided in the future, a supplemental report will be submitted.This product investigation is still in progress.This report will be updated when product investigation is complete.
|
| |
|