| Model Number |
D140 |
| Medical Device Problem Codes |
Signal Artifact/Noise (1036); Over-Sensing (1438)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/24/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during a review of the recorded episodes of this implantable cardioverter defibrillator (icd), an isolated noise oversensing episode was observed.Technical services (ts) noted that the noise was likely from electromagnetic interference during a procedure, since therapy was programmed off for the episode.No adverse patient effects were reported.This device remains in service.
|
| |
|
Additional Manufacturer Narrative
|
|
This product remains implanted and has not been returned to boston scientific, and as a result, laboratory analysis could not be conducted.Based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of known inherent risk of device.
|
| |
|