Cassette (b)(6) was returned and verified to be in use at the time of the event.During investigation testing, a cassette leak was confirmed.Upon disassembly, a deformation of the measurement valve edge was confirmed, consistent with the observed leakage.The deformation was determined to be associated with a manufacturing process.The contract manufacturer was notified of the investigation findings and will continue to monitor the issue.Upon disassembly of returned twiist pump, serial number (b)(6), fluid ingress was confirmed.The observed ingress was consistent with the leak identified in the returned cassette, as well as the reported pump error alarm.---
the current version of the twiist automated insulin delivery (aid) system user guide provides information regarding system alarms and the recommended actions associated with them.Users are also instructed to have a backup insulin therapy available at all times.No additional information relevant to the reported event has been made available to millyard advanced medical products, llc, for further investigation.
|