Review of the system logs confirmed the reported pump error alarms.Upon disassembly of returned twiist pump, serial number (b)(6), a mechanical malfunction was confirmed, consistent with the reported pump error alarms.A specific cause for the observed malfunction could not be conclusively determined.The contract designer was notified of the investigation findings and will continue to monitor the issue.---
the current version of the twiist automated insulin delivery (aid) system user guide provides information regarding system alarms and the recommended actions associated with them.Users are also instructed to have a backup insulin therapy available at all times.No additional information relevant to the reported event has been made available to millyard advanced medical products, llc, for further investigation.
|