• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION INCEPTA? ICD; Implantable Cardioverter Defibrillator (ICD)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION INCEPTA? ICD; Implantable Cardioverter Defibrillator (ICD) Back to Search Results
Model Number E160
Medical Device Problem Codes High impedance (1291); Inappropriate/Inadequate Shock/Stimulation (1574)
Health Effect - Clinical Code Shock from Patient Lead(s) (3162)
Date of Event 01/15/2025
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient with this implantable cardioverter defibrillator (icd) visited the emergency room (er) for evaluation after receiving shocks.Technical services (ts) was called for consult review of device data.The device was interrogated however, due to device being in end of life (eol) battery status, the episodes were unable to be reviewed to determine shocks were appropriate or not because device does not store episodes when in eol status.Ts reviewed the stored daily threshold and impedance measurements trend reports which showed high out of range shock impedances measurements on the right ventricular (rv) lead that appeared to have been gradually increasing since 2018.Subsequently, later on, a generator replacement procedure was performed.This icd was explanted due to normal battery depletion (nbd) and replaced with a new device successfully.The rv lead remains in service.No additional adverse patient effects were reported.
 
Additional Manufacturer Narrative
If pertinent information is provided in the future, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INCEPTA? ICD
Common Device Name
Implantable Cardioverter Defibrillator (ICD)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
Boston Scientific Corporation
cashel road
clonmel E91 T 862
EI   E91 T862
Manufacturer Contact
mark mischke
4100 hamline avenue north
saint paul, MN 55112
+1(612)875-7234
MDR Report Key50601337
Report Number2124215-2026-46718
Device Sequence Number10710117
Product Code LWP
UDI-Device Identifier00802526480751
UDI-Public(01)00802526480751(17)160724
Combination Product (Y/N)N
Initial Reporter StateLA
Initial Reporter CountryUS
PMA/510(K) Number
P960040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2026
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date07/24/2016
Device Model NumberE160
Device Catalogue NumberE160
Device Lot Number111158
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date09/04/2026
Date Device Manufactured07/30/2014
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age63 YR
Patient SexMale
-
-