| Catalog Number |
08H67-01 |
| Medical Device Problem Code |
Low Test Results (2458)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/18/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer observed falsely decreased hemoglobin (hgb) results by the cell dyn ruby analyzer for a 82yrs old female patient.The results provided were:
on (b)(6) 2026 sid (b)(6) hemoglobin: initial=6.44 g/dl /repeated=6.31 g/dl
redrawn and repeated sid (b)(6)=10.1 g/dl.On (b)(6) 2026=9.4 g/dl
on (b)(6) 2026=10.2 g/dl
on (b)(6)2026=11.0 g/dl
on (b)(6) 2026=12.3 g/dl
on (b)(6) 2026=10.7 g/dl
there was no reported impact to patient management.
|
| |
|
Event or Problem Description
|
|
The customer observed falsely decreased hemoglobin (hgb) results by the cell dyn ruby analyzer for a 82yrs old female patient.The results provided were:
on (b)(6)2026 sid (b)(6) hemoglobin: initial=6.44 g/dl /repeated=6.31 g/dl.Redrawn and repeated sid (b)(6) =10.1 g/dl.On (b)(6)2026=9.4 g/dl.On (b)(6)2026=10.2 g/dl.On (b)(6)2026=11.0 g/dl.On (b)(6)2026=12.3 g/dl.On (b)(6)2026=10.7 g/dl.There was no reported impact to patient management.
|
| |
|
Additional Manufacturer Narrative
|
|
An evaluation is in process.A follow-up report will be submitted when the evaluation is complete.Section a patient information: refer to section b5 for complete patient information.All available patient information was included.Additional patient details are not available.
|
| |
|
Additional Manufacturer Narrative
|
|
Data and information provided by the customer were reviewed and support the complaint issue.No instrument troubleshooting has been performed, and sample integrity cannot be ruled out as a contributing factor to the reported result discrepancy due to the difficulties encountered during sample collection.Review of the instrument service history found no subsequent complaints for discrepant patient hemoglobin results.There were no additional service or complaint issues on or around the date this complaint was initiated that may have contributed to this issue.A review of complaints and trending data for cell-dyn ruby analyzer did not identify any similar issues related to the current issue.A review of the manufacturing documentation did not identify any non-conformances and potential non-conformances related to the current issue.Labeling was reviewed and found to be adequate.Based on the available information, no systemic issue or deficiency of the cell-dyn ruby analyzer for serial number (b)(6) was identified.
|
| |
|