Model Number M312
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Pocket Erosion (2013)
Date of Event 08/11/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a connection check was performed, but no response was received from the insertable cardiac monitor (icm).The health care professional (hcp) was unable to feel the device, although the incision site was visible.The issue remained unresolved.A possible device erosion was suspected, and a chest x-ray was recommended to confirm the icm location.Additional information received from the field representative indicates that no infection was noted in the patient, and no further action has been taken at this time.No additional adverse patient effects were reported.This icm remains in service.
 
Event or Problem Description
It was reported that a connection check was performed, but no response was received from the insertable cardiac monitor (icm).The health care professional (hcp) was unable to feel the device, although the incision site was visible.The issue remained unresolved.A possible device erosion was suspected, and a chest x-ray was recommended to confirm the icm location.Additional information received from the field representative indicates that no infection was noted in the patient, and no further action has been taken at this time.No additional adverse patient effects were reported.This icm remains in service.
 
Additional Manufacturer Narrative
This product investigation is still in progress.This report will be updated when product investigation is complete.
 
Additional Manufacturer Narrative
This product has not been returned to boston scientific, and as a result, laboratory analysis could not be conducted.A review of the device history record (dhr) was performed.The review of the dhr identified that there were no process related non-conformances, scrap, or rework performed during the production that could explain the event.The reviews ensure each device meets specification prior to release for use.There is no indication the device manufacturing process contributed to the reported complaint.Based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of no problem detected.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LUX-Dx II+?
Common Device Name
Insertable Cardiac Monitor
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
Boston Scientific Corporation
cashel road
clonmel E91 T 862
EI   E91 T862
Manufacturer Contact
mark mischke
4100 hamline avenue north
saint paul, MN 55112
+1(612)875-7234
MDR Report Key50602106
Report Number2124215-2026-45888
Device Sequence Number10711644
Product Code MXD
UDI-Device Identifier00802526620706
UDI-Public(01)00802526620706(17)270210
Combination Product (Y/N)N
Initial Reporter StateSC
Initial Reporter CountryUS
PMA/510(K) Number
K231328
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/21/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model NumberM312
Device Catalogue NumberM312
Device Lot Number160221
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Supplement Date Received by Manufacturer09/17/2026
Initial Report FDA Received Date09/04/2026
Supplement Report FDA Received Date09/21/2026
Date Device Manufactured08/26/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age62 YR
Patient SexFemale
-
-