• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION INGEVITY? MRI; Implantable Lead

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION INGEVITY? MRI; Implantable Lead Back to Search Results
Model Number 7740
Medical Device Problem Codes Failure to Capture (1081); High impedance (1291); Low impedance (2285)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/29/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the implantable cardioverter defibrillator (icd) and this right atrial (ra) lead exhibited a high out of range pacing impedance.The pacing impedance ranged from 500-600 ohms, then started to dip into 200 ohms, then went to greater than 3000 ohms and has stayed at that measurement.Also, technical services (ts) reviewed presenting electrograms (egm) and suspected there was a loss of capture on this ra lead and the right ventricular (rv) lead.Ts recommended evaluating lead threshold with surface in clinic.Ts confirmed the icd had an appropriate safety margin based on the last in office measurement.This ra lead remains in service.No adverse patient effects were reported.
 
Additional Manufacturer Narrative
If pertinent information is provided in the future, a supplemental report will be submitted.This product investigation is still in progress.This report will be updated when product investigation is complete.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INGEVITY? MRI
Common Device Name
Implantable Lead
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
Boston Scientific Corporation
road 698, lot no. 12
dorado PR 00646 -260
PR   00646-2602
Manufacturer Contact
mark mischke
4100 hamline avenue north
saint paul, MN 55112
+1(612)875-7234
MDR Report Key50602212
Report Number2124215-2026-45921
Device Sequence Number13401754
Product Code NVN
UDI-Device Identifier00802526523427
UDI-Public(01)00802526523427(17)180614
Combination Product (Y/N)Y
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
P150012
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 09/09/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date06/14/2018
Device Model Number7740
Device Catalogue Number7740
Device Lot Number638057
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Supplement Date Received by Manufacturer08/31/2026
Initial Report FDA Received Date09/04/2026
Supplement Report FDA Received Date09/09/2026
Date Device Manufactured06/14/2016
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age46 YR
Patient SexMale
-
-