| Model Number |
D433 |
| Medical Device Problem Codes |
Signal Artifact/Noise (1036); Over-Sensing (1438)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/17/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that this implantable cardioverter defibrillator (icd) exhibited oversensing of electromagnetic interference (emi) noise that resulted in a false signal artifact monitor (sam) event.Minute ventilation (mv) was disabled by the sam event.Technical services (ts) confirmed measurements were in range.Also, ts suspected the patient had another implanted device such as a spinal cord stimulator (scs) or deep brain stimulation (dbs) due to observing persistent noise on the shock electrogram (egm).There was no oversensing noted on the pacing and sensing channels.Ts recommended the health care professional (hcp) ask the patient what they were doing when the noise was observed.This icd remains in service.No adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
This product investigation is still in progress.This report will be updated when product investigation is complete.
|
| |
|