| Model Number |
U125 |
| Medical Device Problem Codes |
Premature Discharge of Battery (1057); Data Problem (3196)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/17/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that this device recorded a code 1003, indicating the voltage was too low for the projected remaining capacity.Additionally, no data was available.Technical services (ts) was consulted and stated that data will need to be transmitted from home or through an in office check.A request was made to have data from this device analyzed.Data analysis identified potential high impedance within the battery.Continued monitoring was recommended.The device will remain implanted, and the patient will be monitored at this time.No adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
This product investigation is still in progress.This report will be updated when product investigation is complete.If pertinent information is provided in the future, a supplemental report will be submitted.
|
| |
|