|
From the information provided, there is no indication that there was any device malfunction, nonconformance, or misuse that contributed to the reported event.Potential adverse events in the instructions for use (ifu) with the tablo system includes, but are not limited to, other, more serious, complications arising from dialysis, such as hemorrhage, air embolism, acidosis, alkalosis or hemolysis, can cause serious patient injury or death.An outset medical, inc.Technical service engineer (tse) has reviewed site device logs with a procedure date of (b)(6) 2026, no related system alarms were found to have occurred during treatment.Additionally, cross functional review of the treatment logs and event description confirmed normal device operation and did not identify any anomalous system activity that could have contributed to the reported event.Based on the available evidence, no device-related failure was identified.A review of production records for this serial number did not note any related manufacturing non-conformances that would contribute to this event.
|