Model Number D152
Medical Device Problem Code Premature Discharge of Battery (1057)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/09/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that this implantable cardioverter defibrillator (icd) emitted beeping tones due to the recorded code 1003, indicating the voltage was too low for the projected remaining capacity.A request was made to have data from this device analyzed.Data analysis showed the battery voltage recently began dropping in an irregular fashion.Device replacement was recommended.As of this time, this icd remains in service.No adverse patient effects were reported.Additional information indicated that this icd was explanted and replaced with the same model.No additional adverse patient effects were reported.
 
Additional Manufacturer Narrative
The device was not returned for analysis; therefore, a technical analysis could not be performed.Good faith effort attempts were made to try and retrieve the device; however, response did not indicate product return.A review of the device history record (dhr) was performed.The review of the dhr identified that there were no process related non-conformances, scrap, or rework performed during the production that could explain the event.The reviews ensure each device meets specification prior to release for use.There is no indication the device manufacturing process contributed to the reported complaint.A risk review was completed and confirmed that the event of premature discharge of battery was defined in the risk documentation.This event type has been accounted for during product risk analysis to support acceptable risk benefit for the product.Review of labeling determined that the complaint situation was listed in the manual.There was no indication in the complaint that the product was not used in accordance to labeling.The manual was unlikely to be the cause of the reported complaint; translation, wording, or graphics does not require further review a similar complaint review was completed using global similar complaint review tool, which reviewed 2233: premature discharge of battery & 2716: message - error code 1003.The review did not identify any emerging trends, meaningful commonalities, or similar complaints requiring escalation.No additional escalation is required.Based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of cause linked to device but unable to trace more specifically because analysis of device data found higher-than-expected power consumption, consistent with premature battery depletion.However, available information was not sufficient to determine probable cause.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DYNAGEN EL ICD DR
Common Device Name
Implantable cardioverter defibrillator
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
Boston Scientific Corporation
cashel road
clonmel E91 T 862
EI   E91 T862
Manufacturer Contact
mark mischke
4100 hamline avenue north
saint paul, MN 55112
+1(612)875-7234
MDR Report Key50602476
Report Number2124215-2026-46238
Device Sequence Number16091890
Product Code LWS
UDI-Device Identifier00802526534324
UDI-Public(01)00802526534324(17)221102
Combination Product (Y/N)N
Initial Reporter StateKY
Initial Reporter CountryUS
PMA/510(K) Number
P960040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2026
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/13/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date11/02/2022
Device Model NumberD152
Device Catalogue NumberD152
Device Lot Number590328
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Supplement Date Received by Manufacturer08/26/2026
09/11/2026
Initial Report FDA Received Date09/04/2026
Supplement Report FDA Received Date09/10/2026
09/14/2026
Date Device Manufactured11/12/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age58 YR
Patient SexMale
-
-