It was reported that this cardiac resynchronization therapy defibrillator (crt-d) system was scheduled for battery replacement.During the procedure, the non boston scientific left ventricular (lv) lead showed no capture in lv only at 7.5 v, however lv capture in was observed in temp brady while in voo mode.It was determined that lv sensing was programmed off due to right atrial (ra) oversensing, and the lv egm appeared flat during lv threshold testing and temp brady.The only difference was that rvs was allowed during the threshold test, which reset the timing.During the battery replacement procedure, three distinct morphologies were observed.Further testing was performed in right ventricular (rv) only pacing, which had similar capture to the lv.None of the lv ring configurations had capture and the cause was not identified.Crt-d explant was completed successfully.No additional adverse patient effects were reported.Product return was requested.
|