Catalog Number NF9100
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code Hemorrhage/Blood Loss/Bleeding (1888)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This complaint has been identified as b braun medical internal complaint number (b)(4).Forty-six unused samples were returned for evaluation.Upon receipt of the returned samples, a visual evaluation was conducted in accordance with the applicable specification.No visual defects were observed during the evaluation.Furthermore, the samples were subjected to leak testing per the applicable specification, with all samples demonstrating acceptable results.Additionally, in order to replicate the reported incident, each sample was punctured multiple times with a needle and subsequently re-tested for leakage.No leakage was observed following multiple punctures.Based on the samples evaluation, the product met internal specifications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SAFELINE®
Common Device Name
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
B. Braun Medical Inc.
824 12th ave
bethlehem PA 18018
Manufacturer (Section G)
B. Braun Medical Inc.
824 12th ave
bethlehem PA 18018
Manufacturer Contact
elena reifenstein
824 12th ave
bethlehem, PA 18018
+1(833)425-1464
MDR Report Key50611607
Report Number2521402-2026-00828
Device Sequence Number10605906
Product Code FPA
UDI-Device Identifier04046964187520
UDI-Public(01)04046964187520(11)230220(10)0061875568
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
K931377
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/25/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberNF9100
Device Lot Number0061875568
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer08/28/2026
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date09/04/2026
Supplement Report FDA Received Date09/25/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/20/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-