| Catalog Number |
191130 |
| Medical Device Problem Code |
Thermal Decomposition of Device (1071)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/21/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Plant investigation: no parts were returned to the manufacturer for physical evaluation.An on-site evaluation was performed by a fresenius field service technician (fst).Field service technician investigation found objective evidence indicating a product problem, thus the complaint is confirmed.A records review was performed on the reported serial number.An investigation of the device manufacturing records was conducted by the manufacturer.There were no non-conformances, or any associated rework identified during the manufacturing process which could be related to the reported event.In addition, the device history record (dhr) review confirmed the results of the in-progress and final quality control (qc) testing met all requirements.The investigation into the cause of the reported problem was able to be confirmed.
|
| |
|
Event or Problem Description
|
|
A fresenius field service technician (fst) was called on-site to service a 2008t hemodialysis (hd) machine that would not power up.The fst found a burnt spot on the motherboard, which caused chain reaction of multiple boards to be damaged.The fst replaced the power control board, main power cable, motherboard, and ui board for power issue.The fst replaced the actuator board for new found issue, act fail error with troubleshooting power issue.The fst replaced the sensor bd fo cond offset reference failure and would not let the fst calibrate the conductivity cells until the sensor board was replaced.The fst replaced the concentrate connector and threaded end because they were fused together.The fst also used a one-year pm kit to replace o-rings, circlip and spacer to complete concentrate connector.There was no patient involvement or patient injury noted.Additional information was requested but was not received to date.
|
| |
|
Event or Problem Description
|
|
A fresenius field service technician (fst) was called on-site to service a 2008t hemodialysis (hd) machine that would not power up.The fst found a burnt spot on the motherboard, which caused chain reaction of multiple boards to be damaged.The fst replaced the power control board, main power cable, motherboard, and ui board for power issue.The fst replaced the actuator board for new found issue, act fail error with troubleshooting power issue.The fst replaced the sensor bd fo cond offset reference failure and would not let the fst calibrate the conductivity cells until the sensor board was replaced.The fst replaced the concentrate connector and threaded end because they were fused together.The fst also used a one-year pm kit to replace o-rings, circlip and spacer to complete concentrate connector.There was no patient involvement or patient injury noted.Additional information was requested but was not received to date.
|
| |
|
Additional Manufacturer Narrative
|
|
Plant investigation: the plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity.
|
| |
|