• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ERIKA DE REYNOSA, S.A. DE C.V. 2008T HD SYS. W/O CDX W/BIBAG BLUE STAR; Dialyzer, High Permeability With Or Without Sealed Dialysate System

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ERIKA DE REYNOSA, S.A. DE C.V. 2008T HD SYS. W/O CDX W/BIBAG BLUE STAR; Dialyzer, High Permeability With Or Without Sealed Dialysate System Back to Search Results
Catalog Number 191130
Medical Device Problem Code Thermal Decomposition of Device (1071)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/21/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Plant investigation: no parts were returned to the manufacturer for physical evaluation.An on-site evaluation was performed by a fresenius field service technician (fst).Field service technician investigation found objective evidence indicating a product problem, thus the complaint is confirmed.A records review was performed on the reported serial number.An investigation of the device manufacturing records was conducted by the manufacturer.There were no non-conformances, or any associated rework identified during the manufacturing process which could be related to the reported event.In addition, the device history record (dhr) review confirmed the results of the in-progress and final quality control (qc) testing met all requirements.The investigation into the cause of the reported problem was able to be confirmed.
 
Event or Problem Description
A fresenius field service technician (fst) was called on-site to service a 2008t hemodialysis (hd) machine that would not power up.The fst found a burnt spot on the motherboard, which caused chain reaction of multiple boards to be damaged.The fst replaced the power control board, main power cable, motherboard, and ui board for power issue.The fst replaced the actuator board for new found issue, act fail error with troubleshooting power issue.The fst replaced the sensor bd fo cond offset reference failure and would not let the fst calibrate the conductivity cells until the sensor board was replaced.The fst replaced the concentrate connector and threaded end because they were fused together.The fst also used a one-year pm kit to replace o-rings, circlip and spacer to complete concentrate connector.There was no patient involvement or patient injury noted.Additional information was requested but was not received to date.
 
Event or Problem Description
A fresenius field service technician (fst) was called on-site to service a 2008t hemodialysis (hd) machine that would not power up.The fst found a burnt spot on the motherboard, which caused chain reaction of multiple boards to be damaged.The fst replaced the power control board, main power cable, motherboard, and ui board for power issue.The fst replaced the actuator board for new found issue, act fail error with troubleshooting power issue.The fst replaced the sensor bd fo cond offset reference failure and would not let the fst calibrate the conductivity cells until the sensor board was replaced.The fst replaced the concentrate connector and threaded end because they were fused together.The fst also used a one-year pm kit to replace o-rings, circlip and spacer to complete concentrate connector.There was no patient involvement or patient injury noted.Additional information was requested but was not received to date.
 
Additional Manufacturer Narrative
Plant investigation: the plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2008T HD SYS. W/O CDX W/BIBAG BLUE STAR
Common Device Name
Dialyzer, High Permeability With Or Without Sealed Dialysate System
Manufacturer (Section D)
ERIKA DE REYNOSA, S.A. DE C.V.
mike allen #1331
parque industrial reynosa
reynosa 88780
MX  88780
Manufacturer (Section G)
ERIKA DE REYNOSA, S.A. DE C.V.
director, quality systems
900 w zaragosa drive suite d
pharr TX 78577
Manufacturer Contact
jessica trujillo
920 winter st
waltham, MA 02451
+1(617)417-5172
MDR Report Key50611914
Report Number0008030665-2026-02390
Device Sequence Number10742515
Product Code KDI
UDI-Device Identifier00840861102112
UDI-Public00840861102112
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
K243237
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date (Section B) 09/18/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Catalogue Number191130
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Device AgeMO
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Supplement Date Received by Manufacturer09/18/2026
Initial Report FDA Received Date09/04/2026
Supplement Report FDA Received Date09/18/2026
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/05/2025
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-