| Model Number |
H749518130 |
| Medical Device Problem Codes |
Leak/Splash (1354); Air/Gas in Device (4062)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/04/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that catheter issue occurred.An opticross imaging catheter was selected for the ultrasound examination of the target lesion.During preparation, the technician noticed that the flushing tubing was leaking, which prevented proper air removal.The procedure was completed with another of same device.The patient remained stable after the procedure.
|
| |
|
Event or Problem Description
|
|
It was reported that catheter issue occurred.An opticross imaging catheter was selected for the ultrasound examination of the target lesion.During preparation, the technician noticed that the flushing tubing was leaking, which prevented proper air removal.The procedure was completed with another of same device.The patient remained stable after the procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation results:
device analysis findings:
the device was not returned for analysis; therefore, a technical analysis could not be performed.Device history record (dhr) review:
review confirmed that the device met all material, assembly, and performance specifications prior to distribution.Investigation conclusion:
this investigation has been assigned an investigation conclusion code of cause not established following a review of the reported events details, available information, and product record review.The device was not returned for analysis, limiting further assessment.Improper handling, device manipulation, or not following instructions may have contributed, but no additional information was available.
|
| |
|