| Model Number |
5101 |
| Medical Device Problem Codes |
High impedance (1291); Communication or Transmission Problem (2896)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/22/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the patient with a rechargeable snm ipg had a red-light error on their remote that would not clear.All high (open) impedances were discovered, and the error could not be cleared in clinic.A successful revision was done on the patient, with both lead and battery replaced.The patient was doing well post procedure.No patient complications were reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Block d2b:
product code - additional product code qon.Block d10:
model number/catalog number: 1201.Serial number: (b)(6).Batch/lot number: (b)(6).Model/catalog description: snm tined lead.Unique identifier (udi) #(b)(4).Block g4:
premarket / 510(k) # - additional premarket/510(k) p190006.H11: investigation details
the device was not returned for analysis; therefore, a technical analysis could not be performed.Good faith effort attempts were made to try and retrieve the device; however, device was discarded.It was confirmed the device met manufacturing specification prior to distribution and there were no manufacturing deviations which could have contributed to the reported event.Based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of caused not established.
|
| |
|