Model Number 429522
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 08/27/2026
Type of Reportable Event Serious Injury
Event or Problem Description
System was explanted due to infection.Should additional information be received, this file will be updated.
 
Additional Manufacturer Narrative
An infection was observed following the implantation of this biotronik device.The sterilization process was investigated.The validated process assures that all sterilization parameters, such as gas concentration, temperature, humidity, etc., are within its specified ranges for each distributed device.Additionally, an analysis of validated microbiological indicators is performed after every sterilization procedure as evidence of successful completion of the sterilization process.Review of the biotronik complaint database did not reveal any changes regarding the trend for this type of incident.In summary, the infection was not device related.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Acticor 7 HF-T QP DF4 IS4 ProMRI
Common Device Name
CRT-D
Manufacturer (Section D)
Biotronik SE & Co. KG
woermannkehre 1
berlin 12359
GM  12359
Manufacturer Contact
6024 jean road
lake oswego, OR 97035
+1(877)245-9800
MDR Report Key50615324
Report Number1028232-2026-04148
Device Sequence Number16135525
Product Code NIK
UDI-Device Identifier04035479156671
UDI-Public(01)04035479156671(17)270930
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P050023
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Device Explanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 09/04/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number429522
Device Catalogue Number429522
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date09/04/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured09/04/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexMale
-
-