| Catalog Number |
RTL160148 |
| Medical Device Problem Code |
Thermal Decomposition of Device (1071)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/27/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A user facility biomedical technician (bmt) reported the mixer would not fill.The bmt was advised to remove panel from the front.The bmt found two burnt wires and was advised to replace the entire assembly.Additional information was requested but was not received to date.
|
| |
|
Event or Problem Description
|
|
A user facility biomedical technician (bmt) reported the mixer would not fill.The bmt was advised to remove panel from the front.The bmt found two burnt wires and was advised to replace the entire assembly.Additional information was requested but was not received to date.
|
| |
|
Additional Manufacturer Narrative
|
|
Plant investigation: the plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity.
|
| |
|
Additional Manufacturer Narrative
|
|
Plant investigation: no parts were returned to the manufacturer for physical evaluation.A records review was performed on the reported serial number.An investigation of the device manufacturing records was conducted by the manufacturer.There were no non-conformances, or any associated rework identified during the manufacturing process which could be related to the reported event.In addition, the device history record (dhr) review confirmed the results of the in-progress and final quality control (qc) testing met all requirements.Complaint investigation found objective evidence indicating a product problem, thus the complaint is confirmed.
|
| |
|