• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Ethicon Inc. Coated VICRYL Plus; Suture, Absorbable, Synthetic

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Ethicon Inc. Coated VICRYL Plus; Suture, Absorbable, Synthetic Back to Search Results
Catalog Number VCP504G
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Impaired Healing (2378)
Date of Event 05/04/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a patient underwent an entropion or ectropion repair and suture was used.The suture was dissolving before the patient's eye was healed.The patient had to be brought back for revision surgery on (b)(6) 2026 and a different type of suture was used.No further issues were noted.Patients healed appropriately.Device is available for return.Additional information was requested.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This is a combination product, and the event has been reviewed for both the suture and the triclosan.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.A manufacturing record evaluation was performed for the finished device lot number, and no related non-conformances were identified.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent: a photo with the following 3 products was sent: vicryl plus 4-0/ vcp504g lot # 10cp4a.Vicryl plus 5-0/ vcp493g lot # 10cq7q.Vicryl 4-0/ j494h lot # xlbbpqx0.Which product is involved in the event for each patient? is vcp493g 10cq7q vicryl plus 5-0 related to this event? if yes, please specify the issue.Is j494h xlbbpqx0 vicryl 4-0 related to this event? if yes, please specify the issue.For each patient please provide the following information: please provide a timeline with details of when the issues occurred.Please provide the patient's demographic information including age, gender, weight, bmi at the time of index procedure.Date and name of index surgical procedure? the diagnosis and indication for the index surgical procedure? were any concomitant procedures performed? on what tissue was the suture used? what was the tissue condition (normal, thin, calcified, fragile, diseased)? for the original intended closure, how were all layers closed? how was the suture placed (interrupted or continuous)? how was the suture originally tied (multiple knots, square knot, etc.)? please describe any medical/surgical intervention required for this suture event including dates and results.Did the operating surgeon observe any suture deficiency or anomaly before, during, after the suture placement or during any re-operation? what symptoms did the patient experience following the index surgical procedure? onset date? did the patient have a wound dehiscence? did the suture break? if so, where was the break noted (termination, middle, end)? onset date/time of early dissolving/possible dehiscence? (# post op days) please describe the appearance of the suture during the second procedure.Were there any precipitating patient stress factors that led to the suture breakage? are patient photos available? other relevant patient history/concomitant medications? what is the physician¿s opinion as to the etiology of or contributing factors to this event? what is the patient's current status? will product be returned? if so, please provide the return date and tracking information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Coated VICRYL Plus
Common Device Name
Suture, Absorbable, Synthetic
Manufacturer (Section D)
Ethicon Inc.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
Ethicon Inc.-Juarez
calle durango #2751
ciudad juarez 32575
MX   32575
Manufacturer Contact
orla o'mahony
1000 route 202
raritan, NJ 08869
+1(303)552-6892
MDR Report Key50615893
Report Number2210968-2026-08982
Device Sequence Number13404299
Product Code GAM
UDI-Device Identifier10705031053922
UDI-Public(01)10705031053922(17)301130(10)10CP4A
Combination Product (Y/N)Y
Initial Reporter StateKY
Initial Reporter CountryUS
PMA/510(K) Number
K181652
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Distributor
Type of Report Initial
Report Date (Section B) 09/04/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberVCP504G
Device Lot Number10CP4A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date09/04/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured12/13/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-