It was reported that during use, a message from the icu.The patient experienced urinary leakage, and upon examination, the foley catheter balloon was found to be ruptured.
|
|
The reported event was inconclusive because no sample was returned.It is unknown whether the device had met relevant specifications.The product was used for urological care.It was unknown whether the product had caused the reported failure.A potential root cause for this failure mode could be due to short latex dwell time, latex dip speed out too slow causing thin sac.Labeling review not required.For the affected date code, the rejected units were recorded as in process scrap and will be disposed of after inspection.No non conformities or discrepancies were identified in any of the dhrs.This report references a us equivalent device.The us udi equivalent for this product number is used.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|