| Model Number |
G150 |
| Medical Device Problem Codes |
Failure to Read Input Signal (1581); Under-Sensing (1661)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/13/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the patient with this cardiac resynchronization therapy defibrillator (crt-d) was evaluated on an in-clinic visit.An issue with the non-boston scientific left ventricular (lv) and right atrial (ra) leads was identified, and the device was subsequently reprogrammed for right ventricular (rv) only.Technical services (ts) reviewed the presenting electrogram (egm) and determined that the device was undersensing p waves.Ts recommended reprogramming options.No adverse patient effects were reported.This crt-p remains in service.
|
| |
|
Additional Manufacturer Narrative
|
|
The local area sales representative was contacted for additional information.At this time, no further information is available.Should additional information become available this report will be updated.
|
| |
|