Model Number G150
Medical Device Problem Codes Failure to Read Input Signal (1581); Under-Sensing (1661)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/13/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient with this cardiac resynchronization therapy defibrillator (crt-d) was evaluated on an in-clinic visit.An issue with the non-boston scientific left ventricular (lv) and right atrial (ra) leads was identified, and the device was subsequently reprogrammed for right ventricular (rv) only.Technical services (ts) reviewed the presenting electrogram (egm) and determined that the device was undersensing p waves.Ts recommended reprogramming options.No adverse patient effects were reported.This crt-p remains in service.
 
Additional Manufacturer Narrative
The local area sales representative was contacted for additional information.At this time, no further information is available.Should additional information become available this report will be updated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DYNAGEN? CRT-D
Common Device Name
cardiac resynchronization therapy defibrillator (CRT-D)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
Boston Scientific Corporation
cashel road
clonmel E91 T 862
EI   E91 T862
Manufacturer Contact
mark mischke
4100 hamline avenue north
saint paul, MN 55112
+1(612)875-7234
MDR Report Key50616744
Report Number2124215-2026-47405
Device Sequence Number10713412
Product Code NIK
UDI-Device Identifier00802526534614
UDI-Public(01)00802526534614(17)190406
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P010012
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 09/04/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date04/06/2019
Device Model NumberG150
Device Catalogue NumberG150
Device Lot Number486186
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date09/04/2026
Date Device Manufactured04/18/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age73 YR