Model Number 471400-12
Medical Device Problem Code Thermal Decomposition of Device (1071)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/12/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Intuitive surgical inc.(isi), has not received the da vinci product with an alleged issue to perform failure analysis.A review of the provided image was performed by an intuitive surgical, inc.(isi) failure analysis engineer (fae).The following additional information was provided: the images show a long bipolar grasper with thermal damage between the grips.
 
Event or Problem Description
It was reported that during a da vinci-assisted surgical procedure, the long bipolar grasper instrument stopped coagulating and started to smell burnt.The ceramic component at the tip of the instrument appears to have overheated.The procedure was completed with no reported injury.Isi followed up with the initial reporter and obtained the following additional information: no arcing was observed.The non-energy instrument did not come into contact with or collide with another activated energy instrument during the procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EndoWrist
Common Device Name
BIPOLAR GRASPER, LONG
Manufacturer (Section D)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA 95051
Manufacturer (Section G)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA 95051
Manufacturer Contact
izabel nielson
3410 central expressway
santa clara, CA 95051
+1(408)523-2100
MDR Report Key50618888
Report Number2955842-2026-37899
Device Sequence Number10500049
Product Code NAY
UDI-Device Identifier00886874121528
UDI-Public(01)00886874121528(11)260122(10)K12260122
Combination Product (Y/N)N
Initial Reporter CountryRO
PMA/510(K) Number
K214095
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign
Remedial Action Other
Type of Report Initial
Report Date (Section B) 09/03/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number471400-12
Device Catalogue Number471400
Device Lot NumberK12260122 0034
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date09/04/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured01/22/2026
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
da Vinci Instruments and Accessories.
Patient SexUnknown
-
-