| Model Number |
M00558880 |
| Medical Device Problem Codes |
Entrapment of Device (1212); Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/11/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that a cre rx biliary dilatation balloon was used in the esophagus, gastric, and duodenum during an esophagogastroduodenoscopy (egd) procedure to treat stricture performed on (b)(6) 2026.During the procedure, the black outer catheter was detached and left in the scope channel.Subsequently, it was pushed into the patient and was successfully retrieved using a rescue forceps.The device was removed and the procedure was completed cre rx biliary dilatation balloon.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.
|
| |
|
Additional Manufacturer Narrative
|
|
Block d4, h4:
detailed product information was not provided to bsc.Based on the nature of the information provided to bsc, it is not possible to perform a good faith effort to obtain additional information.Because the product is unknown at this time, we are unable to provide the complete unique identifier (udi) # and other product specific information.If additional details become available, a supplemental report will be submitted.Block h2:
block h1 (type of reportable event) has been corrected.Block h6:
imdrf device code a0501 captures the reportable event of balloon rx tunnel detachment.Imdrf impact code f2301 is being used to capture the reportable event of additional device required (rescue forceps).Block h11: investigation results:
the device was not returned for analysis and was reported to be disposed; therefore, a technical analysis could not be performed.Based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of cause not established.A review of the manufacturing documentation for this device was unable to be performed as the lot number is unknown.However, a ship history review was performed to identify the most probable lots, and a manufacturing review of the most probable lots did not identify any anomalies or deviations that could have contributed to the event
a similar complaint review was completed.There were no emerging trends identified that warrant further action.
|
| |
|