Model Number M00558880
Medical Device Problem Codes Entrapment of Device (1212); Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/11/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported to boston scientific corporation that a cre rx biliary dilatation balloon was used in the esophagus, gastric, and duodenum during an esophagogastroduodenoscopy (egd) procedure to treat stricture performed on (b)(6) 2026.During the procedure, the black outer catheter was detached and left in the scope channel.Subsequently, it was pushed into the patient and was successfully retrieved using a rescue forceps.The device was removed and the procedure was completed cre rx biliary dilatation balloon.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.
 
Additional Manufacturer Narrative
Block d4, h4: detailed product information was not provided to bsc.Based on the nature of the information provided to bsc, it is not possible to perform a good faith effort to obtain additional information.Because the product is unknown at this time, we are unable to provide the complete unique identifier (udi) # and other product specific information.If additional details become available, a supplemental report will be submitted.Block h2: block h1 (type of reportable event) has been corrected.Block h6: imdrf device code a0501 captures the reportable event of balloon rx tunnel detachment.Imdrf impact code f2301 is being used to capture the reportable event of additional device required (rescue forceps).Block h11: investigation results: the device was not returned for analysis and was reported to be disposed; therefore, a technical analysis could not be performed.Based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of cause not established.A review of the manufacturing documentation for this device was unable to be performed as the lot number is unknown.However, a ship history review was performed to identify the most probable lots, and a manufacturing review of the most probable lots did not identify any anomalies or deviations that could have contributed to the event a similar complaint review was completed.There were no emerging trends identified that warrant further action.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CRE RX
Common Device Name
Catheter, Biliary, Diagnostic
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
Boston Scientific Cork Limited
cork business technology park
model farm road
cork T12 Y K88
EI   T12 YK88
Manufacturer Contact
christina lam
300 boston scientific way
marlborough, MA 01752
+1(914)571-6918
MDR Report Key50623626
Report Number3005099803-2026-03644
Device Sequence Number13402763
Product Code FGE
UDI-Device Identifier08714729965435
UDI-Public(01)08714729965435
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
K172520
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/16/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberM00558880
Device Catalogue NumberM00558880
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Supplement Date Received by Manufacturer09/10/2026
Initial Report FDA Received Date09/04/2026
Supplement Report FDA Received Date09/17/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-