| Model Number |
H74939171100410 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/21/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
D2b - pro code (product code): fge, lit.(b)(6) hospital, (b)(6).E1 - initial reporter address 1: (b)(6).E1 - initial reporter phone: (b)(6).G4 - premarket / 510(k) #: k141521, k141597.B5: describe event or problem: updated.Based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of adverse event related to patient condition.
|
| |
|
Event or Problem Description
|
|
It was reported that balloon rupture occurred.The 75% stenosed target lesion was located in the subclavian artery.A 10.0 x 40, 135cm mustang balloon was advanced for dilation.However, it was noted that the balloon ruptured when the pressure reached 12 atmospheres during the procedure.The procedure was completed with another of the same device.There were no patient complications as a result of this event, and the patient was stable post-procedure.It was further reported that the target lesion was moderately tortuous and moderately calcified.The balloon completely removed from the patient's body via guiding sheath after air was removed from the balloon.
|
| |
|