Abbott Medical (CRM-Sylmar) TENDRIL ST; Drug Eluting Permanent Right Ventricular (Rv) Or Right Atrial (Ra) Pacemaker Electrodes Back to Search Results
Model Number 1888TC/46
Medical Device Problem Codes Break (1069); Over-Sensing (1438); Pacing Problem (1439); Low impedance (2285)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/21/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient presented to the hospital for a device changeout procedure.Prior to the procedure, it was noted that the right atrial (ra) lead exhibited oversensing noise resulting in inappropriate mode switch, low pacing impedance.During the procedure, it was noted that both the ra and right ventricular (rv) leads had compromised outer insulation and exposed wire.Both the ra and rv leads were repaired and remained implanted.The patient was in stable condition.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TENDRIL ST
Common Device Name
Drug Eluting Permanent Right Ventricular (Rv) Or Right Atrial (Ra) Pacemaker Electrodes
Manufacturer (Section D)
Abbott Medical (CRM-Sylmar)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
Abbott Medical (CRM-Sylmar)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
navpreet singh
15900 valley view court
sylmar, CA 91342
MDR Report Key50659283
Report Number2017865-2026-18865
Device Sequence Number13301568
Product Code NVN
UDI-Device Identifier05414734501743
UDI-Public(01)05414734501743(10)2129527(17)101022
Combination Product (Y/N)Y
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
P960013
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2008
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/08/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/22/2010
Device Model Number1888TC/46
Device Lot Number0002129527
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/21/2026
Initial Report FDA Received Date09/08/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured11/15/2007
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
Assurity MRI Pacemaker.
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-