| Model Number |
L111 |
| Medical Device Problem Code |
Premature Discharge of Battery (1057)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/24/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that this pacemaker recorded a code 1003, indicating the voltage was too low for the projected remaining capacity.A request was made to have data from this device analyzed.Data analysis identified potential high impedance within the battery.Device replacement was recommended.The device remains in service.No adverse patient effects were reported.Additional information was provided that the patient was contacted regarding a recommended generator replacement based on technical services guidance, and advisory information was provided.The device replacement procedure has been scheduled.No adverse patient effects have been reported.
|
| |
|
Additional Manufacturer Narrative
|
|
This product investigation is still in progress.This report will be updated when product investigation is complete.
|
| |
|
Event or Problem Description
|
|
It was reported that this pacemaker recorded a code 1003, indicating the voltage was too low for the projected remaining capacity.A request was made to have data from this device analyzed.Data analysis identified potential high impedance within the battery.Device replacement was recommended.The device remains in service.No adverse patient effects were reported.Additional information was provided that the patient was contacted regarding a recommended generator replacement based on technical services guidance, and advisory information was provided.The device replacement procedure has been scheduled.No adverse patient effects have been reported.Additional information was received indicating that the device was explanted.The product has not been returned for analysis.No additional adverse patient effects have been reported.
|
| |
|
Additional Manufacturer Narrative
|
|
This product investigation is still in progress.This report will be updated when product investigation is complete.
|
| |
|