| Model Number |
D140 |
| Medical Device Problem Code |
Premature Discharge of Battery (1057)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/27/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that this implantable cardioverter defibrillator (icd) recorded a code 1003, which states that the voltage is too low for projected remaining capacity.A request was made to have data from this device analyzed.Data analysis confirmed that the code appears consistent with premature battery depletion due to hydrogen degradation of the bypass capacitors.Device replacement was recommended.This device remains in service at this time.No adverse patient effects were reported.
|
| |
|
Event or Problem Description
|
|
It was reported that this implantable cardioverter defibrillator (icd) recorded a code 1003, which states that the voltage is too low for projected remaining capacity.A request was made to have data from this device analyzed.Data analysis confirmed that the code appears consistent with premature battery depletion due to hydrogen degradation of the bypass capacitors.Device replacement was recommended.This device remains in service at this time.No adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
This product has not been returned to boston scientific, and as a result, laboratory analysis could not be conducted.A review of the device history record (dhr) was performed.The review of the dhr identified that there were no process related non-conformances, scrap, or rework performed during the production that could explain the event.The reviews ensure each device meets specification prior to release for use.There is no indication the device manufacturing process contributed to the reported complaint.A risk review was completed and confirmed that the event of premature discharge of battery was defined in the risk documentation.This event type has been accounted for during product risk analysis to support acceptable risk benefit for the product.Review of labeling determined that the complaint situation was listed in the manual.There was no indication in the complaint that the product was not used in accordance to labeling.The manual was unlikely to be the cause of the reported complaint; translation, wording, or graphics does not require further review.Based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code cause linked to device but unable to trace more specifically.
|
| |
|
Search Alerts/Recalls
|