• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Abbott Medical (CRM-Sylmar) Aveir¿ Leadless Pacemaker, Right Ventricle

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Abbott Medical (CRM-Sylmar) Aveir¿ Leadless Pacemaker, Right Ventricle Back to Search Results
Model Number LSP112V
Medical Device Problem Codes Stretched (1601); Use of Device Problem (1670); Impedance Problem (2950)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/24/2026
Type of Reportable Event Malfunction
Event or Problem Description
During an initial implant procedure, low current of injury (coi) and varying low impedance values were observed on the leadless pacemaker (lp).The lp was repositioned and the helix became stretched after interacting with a valve.The lp was explanted and replaced to resolve the event.The patient is stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Aveir¿ Leadless Pacemaker, Right Ventricle
Common Device Name
Leadless Pacemaker
Manufacturer (Section D)
Abbott Medical (CRM-Sylmar)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
Abbott Medical (CRM-Sylmar)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
navpreet singh
15900 valley view court
sylmar, CA 91342
MDR Report Key50660192
Report Number2017865-2026-18626
Device Sequence Number10158868
Product Code PNJ
UDI-Device Identifier05415067034472
UDI-Public(01)05415067034472(10)S000115082(17)280717
Combination Product (Y/N)Y
Initial Reporter CountrySZ
PMA/510(K) Number
P150035
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Device Explanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/07/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberLSP112V
Device Lot NumberS000115082
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/24/2026
Initial Report FDA Received Date09/08/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured07/17/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-