Product analysis: analysis was able to confirm the customer¿s comment that the external pulse generator (epg) was unable to turn on due to the main printed circuit board (pcb) being corroded and the intravenous (iv) pole hanger being broken.It was also noted that the upper case, keypad flex, liquid crystal display (lcd), display frame, all display grommets, and the screws for the display frame, case, main pcb, and hanger were contaminated.The main seal and one plug were damaged.All found defective parts were replaced and all other identified issues were resolved.The device passed all final functional tests.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|