| Catalog Number |
04745914001 |
| Medical Device Problem Code |
Low Test Results (2458)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/12/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
There was an allegation of questionable calcium gen.2 results for 1 patient sample on a cobas 6000 c501 module.The initial result was (b)(6), and the repeated result was (b)(6).The questionable result was not reported outside the laboratory.
|
| |
|
Additional Manufacturer Narrative
|
|
The reagent lot number and expiration date were not provided.The field service representative cleaned the sample probe, the cell rinse station tubing, and the water container.He adjusted the gear pump pressure.He performed checks and tests with acceptable results.The investigation determined that the service actions resolved the issue.
|
| |
|