Patient reported "implant rupture", "cysts", "small avascular cysts", "fibroadenoma", "fibroadenomatosis", "shallow folds", "integrity of the endoprosthesis impaired in the projection of the inner quadrant", "heliomas", "silicogranulomas", "capsule is unevenly thickened", double circuit of the endoprosthesis, "axillary lymph nodes on the sides are increased" and "axillary lymph node hyperplasia".Later healthcare professional reported "pain", "changes in shape", "density" and "implant rupture".Healthcare professional later reported capsular contracture baker grade ii.This record is for the left side.The device was explanted.
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.Device evaluation:
it is not possible to determine the lot number or catalog number through the photos provided.Visual analysis of the photographs identified:
rupture: observed, opening on the device but cannot perform an assessment of the opening as no microscopic analysis can be performed.Capsular contracture: unable to observe since it is not related to the device.No additional observations are performed.No further actions are required since no issue in the manufacturing of the device is observed.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: rupture.
|