The customer reported that recently, some of the tubing does not fit correctly and is unable to create the suction needed.Medline quality received 1 ea of item or56a for evaluation, and functional inspection was conducted with the provided sample.No abnormalities have been found, the item works well and creates a normal suction; at this time, the reported concern could not be confirmed.The cause was documented as not confirmed with received sample, and the division will continue to monitor this issue for trending.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|