The customer reported that the gloves were ripping consistently during use, and after one of the cases, the dr.Noticed the glove was torn and a piece was missing that they could not find.Medline conducted a thorough investigation, which included a review of the reported concern, examination of the returned sample, and comparison against the approved product specification measurements on file; thirteen gloves were drawn from the size 8.5 box and tested using visual confirmation, dimensional inspection, donning/doffing test, and grip test.All thirteen gloves passed all testing methods, with no abnormalities observed, and with the provided samples, medline was unable to confirm the reported issue.The root cause could not be identified, and there were no reports of death or serious injury associated with this event.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|