| Catalog Number |
51-109070 |
| Medical Device Problem Code |
Tear, Rip or Hole in Device Packaging (2385)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that damaged sterile packaging and pouch were identified during inspection of circulated stock items.No additional information available for the reported event.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).G2: foreign ¿ japan.Visual examination of the returned cartons confirmed item and lot label identification.Blisters were opened to evaluate the pouch and pre-tear.All devices have creasing and tearing on the pouches at the pre-tear.No other damage was noted during visual evaluation.Pouch seal widths were measured and were less than the acceptable 6mm.Review of the device history records identified no deviations or anomalies during manufacturing.The root cause of the reported event can be attributed to transit damage and a packaging design issue.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor trends.
|
| |
|