• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION AMS 700 CXR; prosthesis, penis, inflatable

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION AMS 700 CXR; prosthesis, penis, inflatable Back to Search Results
Model Number 72404014
Medical Device Problem Codes Mechanical Problem (1384); Material Puncture/Hole (1504)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/04/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient went to the clinic because this inflatable penile prosthesis (ipp) was not working properly.Two weeks later, a surgical procedure was performed, during which a hole in the cylinder was identified.The cause of the cylinder hole was not detailed by the hospital.All the components were removed and replaced.No patient complications were reported.
 
Additional Manufacturer Narrative
Model number/catalog number: 72404310.Serial number: n/a.Batch/lot number: 1000062373.Model/catalog description: pump ms.Unique identifier (udi) #: (b)(4).Model number/catalog number: 72404161.Serial number: n/a.Batch/lot number: 1000167272.Model/catalog description: reservoir 65ml pc.Unique identifier (udi) #: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMS 700 CXR
Common Device Name
prosthesis, penis, inflatable
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
Boston Scientific Corporation
300 boston scientific way
marlborough MA 01752
Manufacturer Contact
farshad fahimi
4100 hamline avenue north
building c
saint paul, MN 55112
MDR Report Key50668391
Report Number2124215-2026-47742
Device Sequence Number10158635
Product Code JCW
UDI-Device Identifier00878953002712
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
N970012
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2026
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/08/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date10/04/2022
Device Model Number72404014
Device Catalogue Number72404014
Device Lot Number0190348004
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/18/2026
Initial Report FDA Received Date09/08/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age67 YR
Patient SexMale
-
-