Abbott Medical (CRM-Sylmar) Unify Assura CRT-D RF HV; Defibrillator, Automatic Implantable Cardioverter, With Cardiac Resynchronization (Crt-D) Back to Search Results
Model Number CD3357-40C
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 08/18/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient presented with infection, endocarditis and tricuspid valve vegetation.The physician explanted the active system ¿ implantable cardioverter defibrillator, right ventricular lead, left ventricular lead, right atrial lead ¿ and the three older inactive leads.A new leadless pacemaker was implanted.The patient was in stable condition.
 
Event or Problem Description
New information received stated that the physician did not make any claim that the infection was related to the abbott product or product-related procedure.
 
Additional Manufacturer Narrative
The device was returned and visual examinations revealed no anomalies.Interrogation and preliminary test results were acceptable.A device history record (dhr) review was performed and review of the sterilization records confirmed normal sterilization cycles for the products.The device met specifications prior to leaving abbott manufacturing facilities.The cause of infection could not be traced to the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Unify Assura CRT-D RF HV
Common Device Name
Defibrillator, Automatic Implantable Cardioverter, With Cardiac Resynchronization (Crt-D)
Manufacturer (Section D)
Abbott Medical (CRM-Sylmar)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
Abbott Medical (CRM-Sylmar)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
navpreet singh
15900 valley view court
sylmar, CA 91342
MDR Report Key50668745
Report Number2017865-2026-18903
Device Sequence Number9513438
Product Code NIK
UDI-Device Identifier05414734508186
UDI-Public(01)05414734508186(10)P000167690(17)250331
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
P030054
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Device Explanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/08/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/31/2025
Device Model NumberCD3357-40C
Device Catalogue NumberCD3357-40C
Device Lot NumberP000167690
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/18/2026
Initial Report FDA Received Date09/08/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured04/11/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age45 YR
Patient SexFemale
-
-