The customer reported that ¿the tips on the syringe keep breaking¿ and that this happened to ¿3 different students and with 3 different kits.¿ medline quality received a photograph of the reported product, and visual assessment of the provided image revealed that the sterile water syringe had broken off into the balloon inflation port.The reported issue was verified by visual assessment of the provided photograph, thus confirming the reported complaint.The cause was documented as ¿comp./part defective, broke,¿ and a retrospective review of the past 6 months preceding the reported issue revealed no additional complaints of broken syringe tips, with no known trends for this product sku.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|