• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Medline Industries, LP MEDLINE; CATHETER,FOLEY,100%SILICONE,24FR,10ML,LF

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Medline Industries, LP MEDLINE; CATHETER,FOLEY,100%SILICONE,24FR,10ML,LF Back to Search Results
Catalog Number DYND11506H
Medical Device Problem Codes Fluid/Blood Leak (1250); Material Split, Cut or Torn (4008)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/30/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that "the tip is plugged off, she said they are tearing a little on the side and causes it to leak badly.".
 
Additional Manufacturer Narrative
The customer reported that the catheters were tearing, causing leakage, and the initial report stated that the tip is plugged off and they are tearing a little on the side.Medline quality did not receive a sample nor photographs of the reported product, and a post-market retrospective review of the past 24 months preceding the reported issue revealed 0 additional complaints of catheter tears.The issue is unconfirmed, as no samples nor photographs were made available to review.The vendor¿s investigation based on the available information could not identify the root cause of this issue; there are currently no known trends for this product and medline quality will continue to monitor for the issue through monthly complaint meetings and quarterly trending analysis.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDLINE
Common Device Name
CATHETER,FOLEY,100%SILICONE,24FR,10ML,LF
Manufacturer (Section D)
Medline Industries, LP
three lakes drive
northfield IL 60093 -275
Manufacturer Contact
daniel soto
three lakes drive
northfield, IL 60093--275
MDR Report Key50669896
Report Number1417592-2026-33738
Device Sequence Number21372002
Product Code EZL
UDI-Device Identifier10080196848829
UDI-Public10080196848829
Combination Product (Y/N)N
Initial Reporter StateSC
Initial Reporter CountryUS
PMA/510(K) Number
K082815
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Type of Report Initial
Report Date (Section B) 09/03/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberDYND11506H
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/30/2025
Initial Report FDA Received Date09/08/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-